Features:
- Self-generating moving-coil output
- Vertical 0° calibration with top connector
- Circular M10×1 mounting base
Bently Nevada
9200-01-01-10-00 converts absolute case motion into a two-wire voltage proportional to vibration velocity without external excitation.
The vertical calibration and M10×1 base of
9200-01-01-10-00 define the required orientation and mechanical attachment for its stated response.
Output from
9200-01-01-10-00 is carried across connector pins A and B, so load, shielding, and polarity must be preserved at the monitor input.
The Bently Nevada 9200-01-01-10-00 two-wire Seismoprobe velocity transducer may still be available for purchase and support from Apter Power beyond End-Of-Life (EOL) by the manufacturer (OEM).
Bently Nevada 9200-01-01-10-00 two-wire Seismoprobe velocity transducer Manuals,Datasheets,Link
Important Notice: Other accessories, manuals, cables, calibration data, software, etc. are not included with this equipment unless listed in the above stock item description. All prices are shown in USD.
Self-Generating Velocity Signal
- Mechanical motion: Internal coil and magnet movement converts case vibration into voltage.
- Velocity relation: The output magnitude follows vibration velocity at the calibrated sensitivity.
- Excitation: No external sensor excitation supply is required.
- Polarity: Pin A becomes positive relative to pin B when case velocity moves toward the connector.
Calibration Angle and Mounting Hardware
- Calibration: The 01 option sets vertical 0° calibration with ±2.5° tolerance.
- Base: The circular base mounts with an M10×1 stud.
- Torque: The specified mounting torque is 5.6 N·m.
- Connector direction: The top connector routes the signal cable away from the mounting base.
Usable Dynamic Range
- Lower limit: The minimum operating frequency is 4.5 Hz (270 cpm).
- Upper response: Typical response extends to 1,000 Hz at +0/-3 dB.
- Load: Sensitivity calibration uses a 10 kΩ load.
- Linearity: Amplitude linearity is specified over the stated velocity range at 100 Hz.
External Cable and Monitor Connection
- Connector: A top-mounted connector is supplied without a cable.
- Cable selection: The external two-wire shielded cable must be selected separately.
- Lead length: The maximum lead length is 305 m.
- Monitor input: Shielding, load, and pin polarity must be maintained through the connected monitor channel.