Features:
- The H-capsule cover-flange gauge-pressure configuration uses the 5 to 500 kPa (20 to 2000 inH2O) capsule, horizontal piping with left-side high pressure and 1/2 NPT female, two entries with blind plug.
- The H-capsule cover-flange gauge-pressure configuration provides 4–20 mA DC with digital communication through HART 5/HART 7 protocol, with digital indicator.
- The H-capsule cover-flange gauge-pressure configuration combines 1/4 NPT female cover-flange pressure ports without separate process connectors, S-code wetted assembly with Hastelloy C-276 diaphragm and 304 SST flat 2-inch pipe bracket.
EJA430E-JHS5J-917DB is the complete ordering identity for the configuration defined by H capsule for lower gauge-pressure spans.
It converts gauge pressure into a control-system signal and supports configuration or diagnostic access through HART 5/HART 7 protocol.
Before commissioning, verify H capsule, overpressure limit or cover-flange connection against the installed equipment and project documentation.
Availability is subject to Apter Power inventory and sourcing confirmation.
Verified resources for Yokogawa EJA430E-JHS5J-917DB:
Important Notice: Other accessories, manuals, cables, calibration data, software, etc. are not included with this equipment unless listed in the above stock item description. All prices are shown in USD.
- The H-capsule cover-flange gauge-pressure configuration measures gauge pressure in process lines, vessels and utility systems handling compatible liquids, gases or steam.
- The H-capsule cover-flange gauge-pressure configuration supports pump discharge, header and equipment-pressure control where the selected pressure assembly is suitable.
- The H-capsule cover-flange gauge-pressure configuration supplies pressure feedback for control, monitoring or protective logic through an engineered instrument loop.
- For control, the documented 4–20 mA DC with digital communication output supplies the primary process variable to an analog control or safety input.
- Damping and output characterization can be configured without changing the physical 5 to 500 kPa (20 to 2000 inH2O) capsule selection.
- In the documented control context, failure-current behavior must be coordinated with the receiving controller's upscale or downscale alarm strategy.
- For updates, back up range, damping, output mode, alarm direction and tag data before exchanging the transmitter electronics.
- When reviewing updates, any host update must retain support for HART 5/HART 7 protocol and the selected diagnostic variables used by the control system.
- After restoration, compare the downloaded configuration with the physical 5 to 500 kPa (20 to 2000 inH2O) capsule and 1/4 NPT female cover-flange pressure ports.
- For temperature, the standard electronics are specified for -40 to 85°C (-40 to 185°F), or -30 to 80°C with LCD, subject to the tighter limit imposed by the selected indicator or approval.
- When reviewing temperature, the process interface follows -40 to 120°C (-40 to 248°F), with fill-fluid and approval options checked for any reduced range.
- Temperature effects are evaluated against the selected 5 to 500 kPa (20 to 2000 inH2O) capsule rather than a neighboring measurement code.